MDR - Annex XiII

Original text

Expert opinion

Procedure for custom-made devices

1. For custom-made devices, the manufacturer or its authorised representative shall draw up a statement containing all of the following information:

  • the name and address of the manufacturer, and of all manufacturing sites,
  • if applicable, the name and address of the authorised representative,
  • data allowing identification of the device in question,
  • a statement that the device is intended for exclusive use by a particular patient or user, identified by name, an acronym or a numerical code,
  • the name of the person who made out the prescription and who is authorised by national law by virtue of their professional qualifications to do so, and, where applicable, the name of the health institution concerned
  • the specific characteristics of the product as indicated by the prescription,
  • a statement that the device in question conforms to the general safety and performance requirements set out in Annex I and, where applicable, indicating which general safety and performance requirements have not been fully met, together with the grounds,
  • where applicable, an indication that the device contains or incorporates a medicinal substance, including a human blood or plasma derivative, or tissues or cells of human origin, or of animal origin as referred to in Regulation (EU) No 722/2012.

Custom-made devices

Custom-made devices are medical devices manufactured for a specific patient upon doctor’s request and/or prescription.
 
Conformity to Annex I of the Medical Device Regulation (MDR) 2017/745 must be stated including any not tested or not met general safety and performance requirements.
 
For custom-made devices representing a combined or combination product, an indication that such a medicinal or biologic product is contained, is required.

 

2. The manufacturer shall undertake to keep available for the competent national authorities documentation that indicates its manufacturing site or sites and allows an understanding to be formed of the design, manufacture and performance of the device, including the expected performance, so as to allow assessment of conformity with the requirements of this Regulation.

For custom-made devices, a Design History File (DHF), Device Master Record (DMR), and Device History Record (DHR) have to be kept on file.

3. The manufacturer shall take all the measures necessary to ensure that the manufacturing process produces devices which are manufactured in accordance with the documentation referred to in Section 2.

For custom-made devices, the manufacturing processes as stated in the respective SOPs need to be followed.

In the case that no SOPs exist, SOPs need to be created prior manufacturing custom-made devices in order to fulfil the requirements of a DHF or DMR, respectively.

 

4. The statement referred to in the introductory part of Section 1 shall be kept for a period of at least 10 years after the device has been placed on the market. In the case of implantable devices, the period shall be at least 15 years. Section 8 of Annex IX shall apply.

The retention period of the documentation for a custom-made device is identical to the retention period of a regular medical device.

5. The manufacturer shall review and document experience gained in the post-production phase, including from PMCF as referred to in Part B of Annex XIV, and implement appropriate means to apply any necessary corrective action, In that context, it shall report in accordance with Article 87(1) to the competent authorities any serious incidents or field safety corrective actions or both as soon as it learns of them.

Reporting and CAPA apply to custom-made devices in the same manner as for regular medical devices.